Type |
SP-500I |
|
Maximum infusion rate |
500ml/h (Different syringes have different maximum rate). |
|
Range of infusion rate |
10ml syringe: 0.1-100ml/h
20ml syringe : 0.1-200ml/h
50ml syringe : 0.1-500ml/h |
|
Bolus rate |
10ml syringe : 100ml/h
20ml syringe : 200ml/h
50ml syringe : 500ml/h |
|
Accumulated infusion volume |
0.1-999.9ml |
|
Accuracy |
¡À3£¥ |
|
Power supply |
AC110-240V, 50/60Hz, 25W. |
|
Battery |
Rechargeable lithium battery, 7.4 V, 1650mAh,compliance with the requirements of IEC 60601-1. |
|
Maximum power consumption |
25W, running more than 5 hours after being fully charged. |
|
Battery charge |
When syringe pump is connected to AC power, the battery will automatically start charging (need8~14hrs). |
|
Fuses |
F2.0AL, 250V. |
|
Displayed information |
Flow rate, accumulated infusion volume, syringe size, battery capacity, bed No., AC power connecting indication |
|
Alarm information |
infusion completion, near completion, occlusion, low battery, syringe abnormal, control abnormal, no AC alarm |
|
Maximum size of the shell |
288¡Á130¡Á122 mm£¨length¡Áwidth¡Áheight£© |
|
Maximum weight |
£¼2.4kg |
|
Classification |
Type BF |
|
Water proof |
IP21 |
|
Electrical Shock Proof |
Battery: Class II; AC Power: Type BF . |
|
Shell material |
ABS plastic |
|
Operation environment |
Environment temperature +5¡æ--+40¡æ, atmosphere pressure 86~106kPa£¬relative humidity 20%~90% |
|
Storage |
Environment temperature -15¡æ- +50¡æ, atmosphere pressure 86~106kPa£¬relative humidity20%~90% |
|
Applicable syringes |
10ml, 20ml and 50ml syringe (use ¡®Double Dove¡¯ syringe to test). |
|
Standard |
Comply with the standards of EN ISO14971:2000 +A1:2003, IEC60601-1:1995+A1+A2, EN IEC 60601-1-2:2001,IEC60601-2-24,1998, EN 980:2003, EN 1041:1998, EN60601-1-4:1996+A1:1999,
EN60601-1-6:2004.
|
|
Operating mode |
Continous operation |