Product type |
SP-500II |
|
Maximum injection rate |
1500ml/h (Different injectors have different maximum rate.) |
|
Range of flow rate |
5ml injector: 0.1-100ml/h
10ml injector: 0.1-200ml/h
20ml injector: 0.1-400ml/h
30ml injector: 0.1-600ml/h
50ml injector: 0.1-1500ml/h |
|
Injection increment |
0.1ml/h |
|
Bolus rate |
5ml injector: 100ml/h
10ml injector: 200ml/h
20ml injector: 400ml/h
30ml injector: 400-600ml/h
50ml injector: 400-1500ml/h |
|
Injection mode selection |
1. Rate mode 2. Time mode 3. Body Weight mode |
|
Rate mode |
Speed: 0.1-1500ml/h (maximum value depends on the injector specifications) |
|
Time mode |
Time:1~2000 minutes
Accumulated injection volume: 0.1-999.9ml |
|
Body Weight mode |
Weight: 0.1~300.0kg
Drug: 0.1~999.9(mg)
Volume: 0.1~999.9ml
Dose: 0.1¡«9999.9(restricted to unit, drug, solution, weight and injector specification).
Unit: mg/kg/h, ug/kg/min |
|
KVO rate |
Adjustable range: 0.1ml/h¡«5.0ml/h (begin the KVO function under the occlusion situation) ; pressing STOP key to cancel the KVO function when it is unnecessary in use. |
|
IP grade |
IP21 |
|
Injection pressure |
The maximum pressure is 0.3Mpa, Occlusion alarm pressure value is 40kPa¡«160kPa when pressure value is minimum, injection rate is 25ml/h, the longest alarm time is 10 minutes. When pressure value is minimum, injection rate is 5ml/h, the longest alarm time is 10 minutes. |
|
Preset volume |
0.1ml-9999.9ml |
|
Preset time |
0~99 hours, 0~59 minutes |
|
Accumulated injection volume |
0.1ml-9999.9ml |
|
Accuracy |
¡À3£¥ |
|
Power supply |
AC 100¡«220V£¬50/60Hz |
|
Battery |
Rechargeable lithium polymer battery, 7.4 V, 1600mAh. |
|
Maximum power consumption |
25VA, running more than 2 hours at the rate of 25ml/h after being fully charged. |
|
Battery charge |
When syringe pump is connected to AC power, the battery will automatically recharge. (about 8¡«14 hours to recharge fully.) |
|
Fuses |
T2AL250V¡« |
|
Displayed information |
Flow rate; accumulated injection volume; injector specification; battery capacity; bed No.; AC power indicator, operation indication, etc. |
|
State indication information |
Stop, inject, bolus, KVO (stop flashing, the other indicators flash by turns.) |
|
Alarm information |
near completion, Over, occlusion, low battery, abnormal 1 (communications failure), abnormal 2 (pump stuck), abnormal 3 (wrong parameter), volume limited alarm, no AC power |
|
Maximum size of the outer shell |
288¡Á130¡Á122 mm£¨length¡Áwidth¡Áheight£©, |
|
Maximum weight |
£¼3.0kg |
|
Classification |
Class¢ò, Type BF |
|
Shell material |
ABS plastic |
|
Operating conditions |
Environment temperature 5¡æ~40¡æ, air pressure 80kPa ~106kPa£¬relative humidity ¡Ü80% |
|
Storage conditions |
Environment temperature -40¡æ~55¡æ, air pressure 50kPa ~106kPa£¬relative humidity ¡Ü95% |
|
Applicable syringes |
5ml¡¢10 ml¡¢20 ml¡¢30ml¡¢50 ml syringes |
|
Electrical safety |
In accordance with the stipulation of GB9706.1-1995 |
|
Applied standards |
According to the stipulation of YZB/Yue 001-2008 |