Used for First-Aid, Ambulance, Emergency situation and patients transport in hospital, it can be used for pediatric and adult patients.
A button operation, it is simple to use with digital control and precise parameters, and contains special self test function to ensure the patient¡¯s safe.
It consists of ventilators, 2L Aluminum Oxygen Cylinder, Breathe Circuit, Carrying Package, Special Bracket for ambulance and pressure reducing valve for large oxygen cylinder, etc.
1. Specifications:
|
Kinds of modes£º C, A/C, SIMV, SIGH, CPAP, PEEP |
|
Parameters |
Range |
Remark |
|
Tidal Volume |
50¡«2000 ml |
Digital Precise Adjustable |
|
Respiratory Rate |
1¡«100 bpm |
Digital Precise Adjustable |
|
O2% |
45¡«100% |
Digital Precise Adjustable |
|
Trigger Level |
-20¡«0 hpa |
Digital Precise Adjustable |
|
Peak Pressure |
0¡«60 hPa |
Digital Precise Adjustable |
|
PEEP |
0¡«20 hPa |
Digital Precise Adjustable |
|
SIGH |
1¡«10 times per 100 breath |
Digital Precise Adjustable |
|
I:E |
MIN: 2:1¡«1:4 |
Automatic Adjustable |
|
Over Pressure Relief |
¡Ü 60 hpa |
Internal |
|
Minute Volume |
¡Ý 18 L/min |
|
2. Monitor
|
Display |
Range |
|
Large LCD Screen Data Display |
Tidal Volume, O2%, Sigh, Respiratory Rate, Flow,
Pressure (Synchronized Trigger, Peak Airway, CPAP, PEEP ) |
|
Large LCD Screen Curve Display |
Airway Pressure Time Waveform Display |
|
Alarm Data Display |
High/Low Airway Pressure Alarm, Lack of Gas/Power Alarm,
Parameter Error |
|
Other Display |
Lock, Charge Indicator, Inspiratory/Expiratory Situation,
DC/AC Power Supply |
3. Self Test, Alarm and Protection:
¢Å. Special Self Test£º
|
Portable ventilator complies with the latest National standards£¬such as YY0600.3-2007£¬ISO9001£¬ISO13485£¬CHINA FDA£¬etc |
|
Circuit Connect Prompt |
Remind the operator to connect the breath circuit correctly to access the using state£» |
|
Sound Alarm Test |
Test sound alarm, to guarantee the reliability using |
|
Gas Pressure Test |
Test gas pressure, remind operator to ensure the gas pressure supply |
|
Internal Battery Capacity Test |
Test the capacity of the internal battery |
|
Circuit Leakage Test |
Test Circuit Leakage, to lest circuit drop off or leak and unexpected situation to happen |
¢Æ. Alarm:
|
portable ventilator contains perfect alarm function and response the machine and patient condition timely and effectively. |
|
High Airway Pressure |
Low Airway Pressure |
|
Low Battery Voltage Alarm |
Low Gas Pressure |
|
Parameter Error |
Silence |
|
High Frequency Alarm |
|
|
¢Ç. Effective safety protection function
|
portable ventilator contains kinds of security protection function, £» |
|
Over High Airway
Pressure Protection£º |
The ventilator contains a pressure relief valve, when the airway pressure is higher than 60 hpa, the pressure relief valve will deflate automatically to prevent hyperventilation or barotraumas |
|
Unexpected shutdown or power cut protection£º |
When the ventilator is using normally, the electricity goes off or shutdown£¬it will stop the mechanical ventilation, and the
spontaneous respiratory will be open at the same time. |
|
Parameter Error Protection: |
When the Respiratory Rate and the Tidal Volume are higher or lower at the same time, it will alarm, and correct the Respiratory Rate and Tidal Volume to common data at the same time. |
|
Lock Protection Function |
Through this function can lock screen, to prevent incorrect operation |
4. Dimensions
|
Items |
Specification |
|
Ventilator Dimension |
L x H x W: 30 *16 *26 cm Weight£º3kg |
|
Carrying Package Dimension |
L x H x W: 46*26*40cm Weight£º8kg£¨Contain Ventilator, Cylinder, etc.£© |
|
Packing Dimension |
L x H x W: 62 *41*24cm Weight£º11kg£¨Contain all the accessories£© |
|
Power Schemes
(European or USA Standard) |
¢ñ£ºInternal Battery can be used for 16 hours |
|
¢ò£ºDC 12V£¬2A or optional external batteries |
|
¢ó£ºAC 110V-220VAC£¬50/60HZ Power£º30W |
|
Gas Schemes |
¢ñ£ºLittle oxygen cylinder£ºEquipped with standard 2L Aluminum
oxygen cylinder |
|
¢ò£ºBig oxygen cylinder on vehicle£ºEquipped with special pressure reducing valve, can be connected with the big oxygen cylinder on vehicle |
|
¢ó£º Oxygen Center£ºEquipped with the oxygen center tube according
to the user¡¯s requirement |
5. Spare Parts List
|
Complete accessories can be used easily during the first-aid and emergency. |
|
Part Name |
Number |
Specification |
|
Ventilator |
1 |
Including the Li-Battery and adapter |
|
Oxygen Cylinder |
1 |
2 L Aluminum£¬including the high pressure tube and
oxygen inlet joint |
|
Carrying Package |
1 |
Including accessories bag |
|
Breath Circuit |
1 |
Including bellows, Exhalation Valve, Mask,
Lung Simulation |
|
Oxygen Regulator |
1 |
Including high pressure tube and oxygen inlet joint
( used for big oxygen cylinder) |
|
Charger£¨Adapter£© |
1 |
Outlet£ºDC19.6V 4.7A |
|
Bracket for ambulance |
1 |
Loading and unloading the ventilator conveniently with using tools |
|
Operational CD |
1 |
Including manual |
| |